Feb 7---GE Healthcare recalls OEC 9900 Elite FLOROSCOPES

Manufacturer: -See Below-

Product: Medicals

Start Date: 2010-02-07  End Date: 2010-03-07


PRODUCT
1) GE OEC 9900 Elite ESP, to produce mobile fluoroscopic images of human anatomy. Recall # Z-0004-2010;

2) GE OEC 9900 Elite GSP Fluoroscopes To produce mobile fluoroscopic images of human anatomy. Recall # Z-0005-2010

CODE
All associated serial numbers

RECALLING FIRM/MANUFACTURER
GE Healthcare, Salt Lake City, UT, by letter beginning October 5, 2009. Firm initiated recall is ongoing.

REASON
A software defect - may result in a false indication and warning on the OEC 9900 Workstation and C-arm display that the X-Ray tube Anode is over-heated. This will also cause the displayed air kerma rate and cumulative air kerma to be inaccurate.

VOLUME OF PRODUCT IN COMMERCE
1,395 units

DISTRIBUTION
Nationwide